2.10 Connection of external devices

761 Compact IC

51

Connect the end of the 6.2141.110 Cable marked with "732/1"

to the remote interface 24 of the 761 Compact IC (see Fig. 2).

2 Tubing connection 766 – injection valve

At the 761 Compact IC, loosen the rotary nipple 31 screwed

onto the interior side of connection 3 (see Fig. 3 and Fig. 4).

Take aspirating tubing 4 completely out of connection 3 and

unscrew from connection "1" of injection valve 32.

Cut PEEK capillary 18 installed at the 766 IC Sample Proces-

sor (see 766 Instructions for Use) to the desired length.

Pull the free end of PEEK capillary tubing 18 through the

opening of connection 3 of the 761 Compact IC and screw it

onto connection "1" of injection valve 32 with the help of a

6.2744.010 PEEK compression fitting.

Retighten rotary nipple 31 on the interior side of connection 3

to fix the capillary.

3 Tubing connection injection valve – waste

Insert 6.2744.020 Coupling (from 761 accessories) into

connection 3 of the 761 Compact IC.

Screw aspirating tubing 4 onto the 6.2744.020 Coupling and

lead it into the waste container.

4 Settings at the 766 IC Sample Processor

For operation with the 761 Compact IC it is recommended

that the following program "761" is used with the 766 IC

Sample Processor:

parameters
method 761
number of samples: Rack

>start sequence
1 CTL:Rm: INIT
>sample sequence
1 SCN:Rm : Wait1

2 MOVE 1 : sample
3 LIFT: 1 : work mm

4 SCN:Rm : Wait1

5 PUMP 1.1 : 120 s

>final sequence
>changer settings
rack number 0
lift rate 1 12 mm/s
shift rate 20
>manual stop
CTL Rmt: **************
CTL RS232:
------------

Method name

Number of samples to be processed
(entire sample rack)

Initialize remote interface

Waiting until 761 Compact IC sends
signal on remote line 3

Move needle to sample position

Place lift with needle to working
position

Waiting until 761 Compact IC sends
signal on remote line 3

Fill sample loop with sample during
120 s

Settings for changer functions

Reaction to manual stop