background image

4 Operation

761 Compact IC

88

Time program

On the

Program

tab of the system settings window a user-defined time

program for instrument control can be entered. This program is started

automatically as defined in the

Start mode

window (see section 4.3.3) ei-

ther at the moment the determination is started (

Start with determination

)

or at the moment the sample is injected (

Start with inject

).

The

Program

tab contains the two following subpages:

Program

Main time program with all program steps.

Remote configuration

Possibility for creation of user-defined remote

commands.

Program
On the

Program

subpage, program steps including time, program in-

struction and parameter can be entered.

Time (1

st

column)

Time at which program instruction is applied.

Entry range:

0.0 ... 999.9 min

If no time is entered, the program instruction is

applied together with the last instruction with

time entry.

Command (2

nd

column)

Program instruction (see List of program in-

structions).
In addition to these predefined instructions,

user-defined remote commands can be en-

tered if activated on the

Remote configuration

tab.

Parameter (3

rd

column)

Parameter for program instruction (see List of

program instructions).

ENABLED

Enable program start (a disabled program is

not started).

<Add>

Add new program instruction.

<Delete>

Delete selected program instruction.

<Verify>

Test the time program (error messages are dis-

played if program is wrong).