Metrohm 761 Compact IC User Manual

Page 69

Advertising
background image

3.3 Calibration

761 Compact IC

59

Click on

<Apply>

. The

Integration

parameters

window is

opened and the chromatogram is reintegrated.

Repeat this procedure for all other integration parameters

until the result satisfies your expectations. Click on

<Apply>

for each parameter change.

Close the

Integration

parameters

window with

<OK>

.

15 Modify peak allocation

Click on

or select

Calibration / Components

from the

Method

menu. The prepared table for the components fluo-

ride, chloride, nitrate, and sulfate is displayed below the

chromatogram. For all peaks which could be allocated un-

ambiguously to a component this table contains the peak

number in the

Peak

column and the retention time in the

Time

column. In the chromatogram itself this retention time is indi-

cated by the cursor (vertical line).

If the

Peak

column contains a

0

the peak must be manually

allocated to a component. In this case, insert the peak num-

ber from the chromatogram into the

Peak

column and click

the

Time

field on this row with the mouse. The corresponding

retention time is then inserted automatically as the new time

for this component.

For each correctly allocated peak click on the retention time

in the

Time

column. This time is automatically shown in the

chromatogram by the cursor. If the retention time is not lo-

cated in the area of the middle of the peak then the value in

the

Time

column should be adapted accordingly.