Metrohm 761 Compact IC User Manual

Page 40

Advertising
background image

2 Installation

761 Compact IC

30

4 Deaerate pump

Make sure that the aspirating tubing 63 for the high-pressure

pump has been immersed into the eluent.

Click the

<On>

button for

IC pump

to switch on the high-

pressure pump.

Use the syringe inserted into connection 8 to aspirate air until

eluent flows into the syringe.

Click the

<Off>

button for

IC pump

to switch off the high-

pressure pump.

Close the rotary knob on purge valve 39 by turning it in a

clockwise direction (see Fig. 3 and Fig. 4).

Remove the syringe from connection 8.

5 Rinse pulsation dampener

Place a beaker beneath the column connection capillary 28.

Click the

<On>

button for

IC pump

to switch on the high-

pressure pump and rinse the pulsation dampener 60 filled

with isopropanol for ca. 10 min with eluent.

Click the

<Off>

button for

IC pump

to switch off the high-

pressure pump.

6 Reduce flow rate

Reset the original flow rate under

Flow

(e.g.

0.5 mL/min

).

Click

<Send to unit>

to send this value to the 761 Compact IC.