Metrohm 761 Compact IC User Manual

Page 97

Advertising
background image

4.3 Systems

761 Compact IC

87

mouse button and selecting the menu item

Choose color / ...

.

Actual values

Display of actual values

Flow, mL/min

Display of flow rate of the high-pressure pump.

Pressure max, MPa

Display of maximum pressure limit for high-

pressure pump.

Pressure min, MPa

Display of minimum pressure limit for high-

pressure pump.

Full scale,

µ

S/cm

Display of selected full scale range.

Remote lines

Display of current remote line settings.

Valve

Injection

valve

<Inject>

Switch injection valve to "INJECT" position.

<Fill>

Switch injection valve to "FILL" position

Suppressor

Suppressor

module

<Step>

Switch the suppressor module to the next posi-

tion. The time since the last switching of the

suppressor module is displayed in the field be-

side the

<Step>

button.

IC pump

High-pressure

pump

<On>

Start high-pressure pump.

<Off>

Stop high-pressure pump.

Peristaltic pump

Peristaltic

pump

<On>

Start peristaltic pump.

<Off>

Stop peristaltic pump.

System startup values

The system startup values are sent and applied

to the 761 Compact IC each time the system is

connected, a determination is started, or the

values are sent manually with

<Send to unit>

.

Flow, mL/min

Startup value for flow rate of the high-pressure

pump.

Entry range:

0.20

...

2.50 mL/min

Pressure max, MPa

Startup value for maximum pressure limit for

high-pressure pump. This limit is controlled

even without connection to the PC.

Entry range:

0.0

...

25.0 MPa

Pressure min, MPa

Startup value for minimum pressure limit for

high-pressure pump. This limit is controlled

even without connection to the PC.

Entry range:

0.0

...

25.0

MPa

Full scale,

µ

S/cm

Startup value for full scale range.

Selection:

50

,

250

,

1000

µ

S/cm

Remote lines

Startup value for remote line settings 1

8.

Selection:

0

,

1