2.6 High-pressure pump

761 Compact IC

25

2.6 High-pressure

pump

In order to avoid damage to the pump it must never be operated dry.

Each time that the pump is switched on always first check that the

eluent supply has been connected up correctly and that sufficient

eluent is present in the eluent bottle.


2.6.1

Removing the transport security screws

In order to prevent the pump drive from being damaged during trans-

port the pump head is fitted with three transport security screws 18 (see

Fig. 2). These transport security screws must be removed before the

high-pressure pump is started up. Also remove the red sticker attached

to the pump head.

In order to avoid damage to the pump head these three security

screws should be attached to the pump head each time that it is to be

transported.

2.6.2

Installing the pulsation dampener

To protect the column material against pressure drops caused by the

injector, the 6.2620.150 Pulsation dampener MF has to be installed

between the high-pressure pump and the injection valve of the 761

Compact IC. Proceed as follows (see Fig. 10):

1 Install pulsation dampener

Position the pulsation dampener 60 in the interior of the 761

Compact IC on the base.

2 Connection to the pump

Unscrew PEEK capillary 35 of coupling 33 and attach it to

connection 62 of the pulsation dampener 60.

3 Connection to injection valve

Unscrew PEEK capillary 26 of coupling 33 and attach it to

connection 61 of the pulsation dampener 60.

The pulsation dampener is filled with isopropanol and must be rinsed

with eluent before connection to a separating column (see section

2.6.4).

The 6.2620.150 Pulsation dampener can be operated in both direc-

tions.