2.7 Precolumns and separating columns

761 Compact IC

33

2.7.3

Precolumn glass cartridges with cartridge holder

The precolumn glass cartridge METROSEP Anion Dual 1 (6.1006.030)

is inserted in the 6.2828.010 Precolumn cartridge holder as follows (see

Fig. 13):

1 Insert cartridge

Remove end fittings 82 from screw caps 83.

Insert precolumn cartridge 85 in sleeve 84 (mark flow direc-

tion of the column cartridge on sleeve, the column cartridge

should always be operated in the same flow direction).

Screw both screw caps 83 loosely to sleeve 84 by hand.

Screw both end fittings 82 into screw caps 83 so that their

capillary ends are seated in the PTFE seals of the cartridge.

Tighten both screw caps 83 by hand.

2 Connect precolumn

Provide the column connection capillary 28 mounted on the

injection valve with a PEEK compression fitting 54 (see sec-

tion 2.3.5) and screw it tightly onto end fitting 82 on the inlet

side of the precolumn.

Cut off as short a piece as possible from PEEK capillary 56

and fit this with two PEEK compression fittings 54 (see sec-

tion 2.3.5).

Screw prepared capillary 56 to end fitting 82 on the other end

of the precolumn.

3 Rinse precolumn

Place a beaker beneath the outlet capillary of the precolumn.

Open software window for manual system control.

If necessary, modify

Flow rate

to the value suited for the

inserted separating column and click on

<Send to unit>

to

send this value to the 761 Compact IC.

Switch on high-pressure pump (

IC pump

) by clicking

<On>

and rinse precolumn with eluent for ca. 10 min.

Switch off high-pressure pump by clicking

<Off>

.

The rinsed precolumn can be connected directly to the separating

column instead of by means of a PEEK capillary. Proceed as follows:

Screw off PEEK capillary from end fitting 82 of the precolumn.

Screw off end fitting 82 from screw cap 83 of the precolumn.

Screw connection piece 86 onto screw cap 83.

Screw screw cap 87 of separating column 81 onto connection

piece 86.

Advertising