Keypad basics, Priming the instrument, Primary menu – Bio-Rad Bio-Plex Pro™ Wash Stations User Manual

Page 20: Secondary menu

Advertising
background image

Basic Operation and Maintenance

20 Bio-Plex Pro and Bio-Plex Pro II Wash Stations

Keypad Basics

All the functions you need to run Bio-Plex programs are easily accessed using the
keypad on your wash station. The system is organized into primary and secondary
menus.

Primary Menu

The instrument menu has the following options:

Primary

Menu

Options

Functions

Secondary Menu Options

Programs

Wash programs used to process assays

Start, Define/Edit, Show, Clear

Settings

Rarely accessed advanced features

Edit Plates, Options, Bubble Sensor

Procedures

Preparation or maintenance processes

Prime, Rinse, Empty Prime Tray,
Vacuum Filtration

Secondary Menu

If you choose OK when any of the primary menu options (Programs, Settings, and
Procedures) are visible on the LED, you will navigate to the secondary menus. You
may Select a choice from the secondary menu, or press Exit to return to the primary
menu.

Priming the Instrument

You should prime both Channel 1 and 2 after installation and if the instrument has not
been used for an extended period of time. Priming is performed to fill the liquid
system of the instrument and to remove all air from the tubes. A priming step must
also be performed when switching to a different wash buffer.

Caution

Do not use your new instrument until you have primed both channels.

Navigate

to other

options

PROGRAM

< >

OK

<

>

F



Select or

Exit

If you get confused about
where you are, press Exit,
until you no longer see Exit
as an option.