Process step: aspirate, Defining new programs – Bio-Rad Bio-Plex Pro™ Wash Stations User Manual

Page 55

Advertising
background image

Defining New Programs

Bio-Plex Pro and Bio-Plex Pro II Wash Stations

55

Process Step: Aspirate

Select required aspirating time (1 - 20 seconds).

Select required aspirating speed (1 - 20 mm/sec).

Select and confirm aspirating step (ASP)
CX = Cycle Number
PX = Program Step Number

Crossw.Asp YES

- + EXIT OK

POS: BOTTOM

- + EXIT OK

TIME: 4 s

- + EXIT OK

H-SPEED10mm/s

- + EXIT OK

CX PX ASP

- + EXIT OK

ASP.RATE 2

- + EXIT OK

Select YES using – or + to perform crosswise
aspiration or select NO to perform normal aspiration.

Press – or + to select the aspiration position
(BOTTOM; OVERFLOW or CUSTOM).

Select required aspirating rate (1 - 3),
where 1 is the weakest and 3 the strongest rate.

After the process step has been defined the initial
display appears.
CX = Cycle Number
PY = Program Step Number incremented by 1
Press OK to define another aspiration step
or press – or + to select the next program step,
Program End or a new cycle.

CX PY ASP

- + EXIT OK

Note
Low residual volumes cannot be guaranteed if not all wells in a strip are
filled with liquid. To prevent cross contamination, the head speed
should be lowered.