Removing the manifold – Bio-Rad Bio-Plex Pro™ Wash Stations User Manual

Page 69

Advertising
background image

Troubleshooting and Error Messages

Bio-Plex Pro and Bio-Plex Pro II Wash Stations

69

Removing the Manifold

WARNING

Prime without liquid to remove liquids from the tubing system.

The manifold should be removed and thoroughly cleaned at least once every
6 months or whenever needles become blocked.

1. Lift the manifold arm and remove the black manifold arm guard plate by

removing the three screws using the Allen key provided.

2. Lift the manifold arm and remove the two screws that attach the manifold to

the instrument using the Allen key provided.

3. Carefully pull the tubing off the connectors on the rear of the manifold and

remove the manifold.

3 screws

2 screws