Removing the manifold – Bio-Rad Bio-Plex Pro™ Wash Stations User Manual
Page 69
Advertising
Troubleshooting and Error Messages
Bio-Plex Pro and Bio-Plex Pro II Wash Stations
69
Removing the Manifold
WARNING
Prime without liquid to remove liquids from the tubing system.
The manifold should be removed and thoroughly cleaned at least once every
6 months or whenever needles become blocked.
1. Lift the manifold arm and remove the black manifold arm guard plate by
removing the three screws using the Allen key provided.
2. Lift the manifold arm and remove the two screws that attach the manifold to
the instrument using the Allen key provided.
3. Carefully pull the tubing off the connectors on the rear of the manifold and
remove the manifold.
3 screws
2 screws