Troubleshooting and Error Messages

Bio-Plex Pro and Bio-Plex Pro II Wash Stations

77

PC (Process Control) System Error

If the Process Control board does not respond, the following error message is
displayed:

Examples of Error Messages:

PC SYS ERROR

PC FUNCT. ERR

PC

ASP

ERROR

Process Control functionality has been inadvertently activated for your system. Your
system is not equipped for process control, so it must be deactivated. The following
directions describe how to check and disable a PC setting.

Remember to exit to the Primary Menu before performing this procedure. (Press Exit
until the word EXIT disappears.)

Turn instrument off and back on for change to take effect. If the error continues,
call the service technician.

No Aspiration at Vacuum Filtration Error

The following error message appears on the display:

NO ASP AT VACFIL
OK

Press OK to remove the error message and to return to the standby menu.