Complete instrument keypad functions – Bio-Rad Bio-Plex Pro™ Wash Stations User Manual

Page 43

Advertising
background image

Detailed Operating Instructions

Bio-Plex Pro and Bio-Plex Pro II Wash Stations

43

Complete Instrument Keypad Functions

The instrument keypad is used to run and manage wash programs, as well as
vacuum filtration programs, to define plate parameters, adjust certain instrument
settings, and perform instrument procedures (rinse, prime, and empty prime tray).

The following features are available:

Preloaded programs for processing Bio-Plex and Bio-Plex Pro assays (also
appropriate for other standard and magnetic xMAP assays)

20 user definable washing programs, stored under program positions 1 to 20

Up to 60 processing steps per program; each processing step can be identical or
different from the previous step

Adjustable soak time (1 second to 60 minutes and 59 seconds)

Variable shaking settings:

o

High: Linear shaking with a shaking frequency of 25 Hz and a shaking

amplitude of 1 mm

o

Medium: Linear shaking with a shaking frequency of 10 Hz and a

shaking amplitude of 2 mm

o

Low: Linear shaking with a shaking frequency of 5 Hz and a shaking

amplitude of 3 mm

Adjustable dispense rate (drip mode to 500 µl/s)

Adjustable aspirating speed (1 to 3)

Adjustable bottom positions (bottom, custom, overflow)

Two rinse modes (Rinse Day and Rinse Night) can be used to select how the
instrument is rinsed before it is left to stand or switched off

Automatic microplate centering, before starting any wash program

Programmable strip selection, before starting the washing procedure

Advertising