Troubleshooting and Error Messages

Bio-Plex Pro and Bio-Plex Pro II Wash Stations

75

Troubleshooting and
Error Messages

Errors Without Messages

The following errors will not yield an error message from the instrument’s firmware:

Error Description

Possible Causes

See

Display and LED dark

Fuse is defective.

Contact your local Service

Representative

No or incorrect dispensing

Dispensing needles blocked.

Cleaning the Manifold on page 62

No or incorrect aspirating

Aspirating needles blocked.

Cleaning the Manifold on page 62

Error Messages

Plate Error

If the plate sensor does not recognize the plate or no plate is inserted on the plate
carrier, the following message is displayed:

NO PLATE
OK

Press OK and insert the plate correctly on the plate carrier

Program Too Big Error

If the program has been defined with more than 60 process steps (each cycle is
worth 2 steps), the following message is displayed:

PROGR: TOO BIG
OK

Press OK and define a program with less than 60 process steps.

Transport Error

If the instrument cannot move the plate carrier, the following message is displayed:

INIT ERROR
OK

Press OK to remove the error message and the instrument returns to the standby
mode.

Check that the plate carrier is correctly inserted and that the plate carrier system is
clear.