Defining New Programs

50 Bio-Plex Pro and Bio-Plex Pro II Wash Stations

Aspirating Modes

To improve wash efficiency and reduce the residual volume, the manifold must be
positioned so that the aspirating needles are correctly positioned in the wells for
round-bottom, V-shaped bottom, or flat-bottom well microplates.

Normal AspirationMode

For round-bottom or V-shaped bottom wells, the aspirating needles are placed in the
middle of the wells. Only one aspiration position can be selected.

Round Bottom Wells

'V' Shaped Wells

Normal Aspirating Positions

Crosswise Aspiration Mode

With flat bottom wells, the instrument can perform crosswise aspiration using two
aspiration positions per well.
The aspirating needles are set at two positions on the bottom of the wells
(front edge and back edge).

Aspirate Y Offset 1

Aspirate Y Offset 2

Front edge

Back edge

Crosswise Aspiration for Flat Bottom Wells