Process step: soak, Process step: user prompt, Process step: vacuum filtration – Bio-Rad Bio-Plex Pro™ Wash Stations User Manual
Page 59: Defining new programs
Defining New Programs
Bio-Plex Pro and Bio-Plex Pro II Wash Stations
59
Process Step: Soak
Select required soaking time using -/+ keys
(0 – 60 min; 1 - 59 seconds )
Press Chg. to select minutes or seconds.
Select and confirm soak step (SOAK).
CX = Cycle Number
PX = Program Step Number
CX PX SOAK
- + EXIT OK
TIME: 0 min 10 s
- + Chg. OK
SHAKE OFF
- + EXIT OK
The shaking speed can be selected
(OFF, LOW, MEDIUM or HIGH).
CX PY SOAK
- + EXIT OK
After the process step has been defined the initial
display appears.
CX = Cycle Number
PY = Program Step Number incremented by 1
Press OK to define another soak step
or press – or + to select the next program step,
Program End or a new cycle.
Process Step: User Prompt
Select and confirm User Prompt strip on
required position in current program.
CX PX USER PR.
- + EXIT OK
Process Step: Vacuum Filtration
Select required vacuum pressure using the - /+ keys.
Select and confirm Vacuum Filtration step (VAC).
CX = Cycle Number
PX = Program Step Number
CX PX VAC
- + EXIT OK
VAC. -100.0 mBar
- + EXIT OK
Select required soaking time using the - /+ keys
(0 – 60 min; 1 - 59 seconds )
Press Chg. to select minutes or seconds.
TIME: 0 min 10 s
- + Chg. OK
CX PX ASP
- + EXIT OK
After the process step has been defined the initial
display appears.
CX = Cycle Number
PY = Program Step Number incremented by 1
Press OK to define another vacuum filtration step
or press – or + to select the next program step,
Program End or a new cycle.