Defining New Programs

Bio-Plex Pro and Bio-Plex Pro II Wash Stations

59

Process Step: Soak

Select required soaking time using -/+ keys
(0 – 60 min; 1 - 59 seconds )
Press Chg. to select minutes or seconds.

Select and confirm soak step (SOAK).
CX = Cycle Number
PX = Program Step Number

CX PX SOAK

- + EXIT OK

TIME: 0 min 10 s

- + Chg. OK

SHAKE OFF

- + EXIT OK

The shaking speed can be selected
(OFF, LOW, MEDIUM or HIGH).

CX PY SOAK

- + EXIT OK

After the process step has been defined the initial
display appears.
CX = Cycle Number
PY = Program Step Number incremented by 1
Press OK to define another soak step
or press – or + to select the next program step,
Program End or a new cycle.

Process Step: User Prompt

Select and confirm User Prompt strip on
required position in current program.

CX PX USER PR.

- + EXIT OK

Process Step: Vacuum Filtration

Select required vacuum pressure using the - /+ keys.

Select and confirm Vacuum Filtration step (VAC).
CX = Cycle Number
PX = Program Step Number

CX PX VAC

- + EXIT OK

VAC. -100.0 mBar

- + EXIT OK

Select required soaking time using the - /+ keys
(0 – 60 min; 1 - 59 seconds )
Press Chg. to select minutes or seconds.

TIME: 0 min 10 s

- + Chg. OK

CX PX ASP

- + EXIT OK

After the process step has been defined the initial
display appears.
CX = Cycle Number
PY = Program Step Number incremented by 1
Press OK to define another vacuum filtration step