Basic Operation and Maintenance

26 Bio-Plex Pro and Bio-Plex Pro II Wash Stations

Rinse Night

If the instrument will be left to stand for a longer period of time (such as overnight),
perform the Rinse Night procedure using these menu choices:

When the rinsing procedure is completed the following message is displayed:

RINSE NIGHT
STOP END

The manifold remains in the prime tray until END is pressed. Press STOP to abort
the procedure (prime tray will not be aspirated).

Instrument state at the end of the wash

Considerations for leaving the instrument on or off after a rinse procedure.

• Leave the instrument on if you want to soak the manifold overnight

(Recommended during periods of heavy use)

• Shut the instrument off if you do not wish to soak the manifold (Rinse Night

only)

Navigate to PROCEDURE in the primary
menu and press OK.

Use the keys under the < and > symbols to access the
PROCEDURE menu if it is not already displayed

Select and press OK for RINSE

Select and press OK for RINSE NIGHT

Press OK when presented with the option to the left

If presented with TIME you can use the default settings here as well. Or navigate to
VOLUME using < and > .

Press OK when presented with the option to the left

Or use the default time settings

Select CHANNEL 2 (using the choice keys <, >)

At the end of the day always rinse with CHANNEL 2, distilled
H

2

O or equivalent.

Instrument will perform rinse procedure.

Advertising