Defining new programs, Introduction, Program menu – Bio-Rad Bio-Plex Pro™ Wash Stations User Manual

Page 53

Advertising
background image

Defining New Programs

Bio-Plex Pro and Bio-Plex Pro II Wash Stations

53

Defining New Programs

Introduction

The following program items are available:

CYCLE:

The number of cycles the subsequent processing steps
should be performed

ASP:

An aspirate step removes liquid from the wells

DISP:

A dispense step fills the wells with liquid

WASH:

Liquid is dispensed and aspirated simultaneously
creating a circular flow with a maximum volume of 3000
µl in one wash step for increased wash efficiency

SOAK:

During a soak step the liquid remains in the wells for the
set time (with or without shaking)

VAC:

For collecting desired substances by vacuum filtration
using filtration plates

USER PROMPT:

User interaction is requested

FINAL ASPIRATE:

An aspirate step at the end of a program

END PROGRAM:

The program finishes

Program Menu

The Program menu has the following options:

Start

Run a defined wash program

Define/Edit

Define or edit a program on the instrument

Show

View the parameters of a defined program

Clear

Clear a wash program from the instrument’s menu

The Program menu has the following structure:

START

< > EXIT OK

PROGRAM

< > OK

DEFINE EDIT

< > EXIT OK

SHOW

< > EXIT OK

CLEAR

< > EXIT OK

Advertising