Basic Operation and Maintenance

Bio-Plex Pro and Bio-Plex Pro II Wash Stations

21

Perform priming for both channels using the following procedure:

Repeat for channel 2.

Failure to Prime

If you have a new instrument or it has not been used for a long time the pump may
fail to draw fluid from the liquid bottles. This problem is easily solved.

1. Disconnect the L1 tube from the bottle (leaving it connected to the

instrument).

2. Hold the free end of the tube high and pipette 3 - 4 ml of deionized

water to the tube while performing the prime procedure in channel 1
as described above.

Alternately you may carefully use a syringe to inject 1 - 2 ml of water directly into the
L1 port of the instrument, reconnect the tubing, and run a prime procedure.

PRIME

< > OK

CHANNEL 1

- + EXIT OK

WASTE BOTTLE OK?
< > YES

PRIME SOLUTION OK?

< > YES

Confirm rinsing solution and tube in bottle

Select and press OK for PRIME

Select CHANNEL 1 using the choice keys (< , >)

Select OK and the prime procedure starts

Confirm waste bottle is connected and
sufficiently empty

PRIME

< > OK

PROGRAM

< > OK

Instrument will perform the prime step and will
then be ready to perform wash protocols

VOLUME 30 mL
+ - OK

VOLUME

< > OK

Press OK if you see VOLUME. If you see
TIME, scroll down to VOLUME, then press OK

PROCEDURE

< > OK

Navigate to the Procedure in the primary
menu and press OK

Press OK