Bio-Rad Bio-Plex Pro™ Wash Stations User Manual
Page 35
About the Instrument
Bio-Plex Pro and Bio-Plex Pro II Wash Stations
35
PARAMETERS CHARACTERISTICS
Carry over
between wells
<= 1 ppm
The carry over between wells <= 1 ppm cannot be guaranteed if not all wells
are filled with liquid.
Vacuum Filtration
Vacuum range (Bio-
Plex Pro II Only)
-50 to -150 mBar absolute pressure
Power
Supply
Auto-sensing
100 - 120 V or 220 - 240 V, 50/60 Hz
Consumption
< 65 VA
Physical
Outside dimensions
(without Vacuum
Filtration)
Width: 275 mm, Depth: 366 mm, Height: 180.5 mm
(Width: 10.8 in., Depth: 14.4 in., Height: 7.1 in.)
Greater depth and height in service position
Weight
(without Vacuum
Filtration)
6.6 kg
Environmental
Ambient temperature
Operation
15°C to 35°C (59°F to 95°F)
Storage
-20°C to 60°C (-4°F to 140°F)
Relative Humidity
Operation
20 % to 80 %
non-condensing
Storage
10 % to 80 % non-condensing
Others
Overvoltage category
II
Pollution degree
2
Method of disposal
Electronic waste
(infectious waste)