Instrument disinfection, Disinfection solutions – Bio-Rad Bio-Plex Pro™ Wash Stations User Manual

Page 71

Advertising
background image

Troubleshooting and Error Messages

Bio-Plex Pro and Bio-Plex Pro II Wash Stations

71

Instrument Disinfection

WARNING

The disinfection procedure should be performed according to national,
regional, and local regulations.

WARNING

All parts of the instrument that come into contact with potentially
infectious material must be treated as potentially infectious.

It is advisable to adhere to applicable safety precautions, (including the
wearing of powder-free gloves, safety glasses, and protective clothing)
to avoid potential infectious disease contamination when performing the
disinfection procedure.

It is very important that the instrument is thoroughly disinfected before it is removed
from the laboratory or any servicing is performed on it.

Before the instrument is returned to the service center for service or repair, it must be
disinfected and a disinfection certificate completed by the operating authority. If a
disinfection certificate is not supplied, the instrument may not be accepted by the
service center or it may be held by the customs authorities.

Disinfection Solutions

Use the following disinfection solutions for the disinfection procedure:

Decon 90 (Decon Laboratories Limited)

Decon Neutracon (Decon Laboratories Limited)

Microcide SQ (Global Biotechnologies)

• Ethanol

WARNING

Risk of fire and explosion!

Ethanol is flammable and when improperly handled can lead to
explosions. Proper laboratory safety precautions must be observed.