3-3 After Use

z

Avoid mechanical shock to the balance.

z

Do not disassemble the balance. Contact the local A&D dealer if the balance needs service or

repair.

z

Do not use organic solvents to clean the balance. Clean the balance with a lint free cloth that

is moistened with warm water and a mild detergent.

z

Protect the internal parts of the balance from liquid spills and excessive dust.

3-4 Power Supply

z

Do not remove the AC adapter while the internal mass is in motion, for example, right after the

AC adapter is connected, or during calibration using the internal mass.

If the AC adapter is removed under the conditions described above, the internal mass will be

left unsecured, that may cause mechanical damage when the balance is moved.
Before removing the AC adapter, press the ON:OFF key and confirm that zero is displayed.

z

When the AC adapter is connected, the balance is in the standby mode if the standby indicator

is on (refer to “4. DISPLAY SYMBOLS AND KEY OPERATION”). This is a normal state and
does not harm the balance. For accurate weighing, warm up the balance for at least 30
minutes before use.