3 details of the function table – Rice Lake MC Series Mass Comparator, A&D Weighing - MC-6100 User Manual

Page 31

Advertising
background image

29

9-3 Details of the Function Table

Class

Item and Parameter

Description

0

1

Cond

Condition

2

Fast response rate, sensitive value FAST
MID.
Slow response rate, stable value SLOW

Can be changed by response

adjustment. With “hold 1”,

sets the averaging time.

0

1

5t-b

Stability band width

2

Stable range is ±1 digit

Stable range is ±3 digits

The stabilization indicator illuminates

when the display fluctuation is

within the range. With “hold 1”,

sets the averaging time.

0

OFF

Hold

Hold function

1 ON

Holds the display when stable

in animal mode. With “Hold 1”,

ANIMAL turns on.

0 OFF

1

Normal

2 Strong

trc

Zero tracking

3 Very strong

Keeps zero display by tracking

zero drift.

0

5 times/second

5pd

Display refresh rate

1 10 times/second

Period to refresh the display

0

Point (.)

pnt

Decimal point

1 Comma (,)

Decimal point format

0

OFF

p-on

Auto display-ON

1 ON

Turns on the weighing mode

display when the AC adapter

is plugged in.

0

OFF

poff

Auto display-OFF

1 ON (10 minutes)

Turns off the display after 10

minutes of inactivity.

0

OFF

g5i

Capacity indicator

1 ON

Capacity indicator

Zero: 0%,

Maximum capacity: 100%

0

Not used

f1l

Filter

1 Used (when the balance is used as a mass comparator)

0

Sets the display to zero.

p-tr

Zero upon power-on

1

Does not set the display to zero. Displays the previous value.

With this setting, do not perform re-zero operations frequently.

Refer to “Zero upon power-on” on page 33.

0

Performs zero and span calibration

ba5fnc

Environment

Display

5pn

Span calibration

1

Performs span calibration only.

Span calibration using the internal mass is possible with a tare

on the weighing pan.

Refer to “Span calibration” on page 33.

Cl adj

Clock

Refer to the GX series instruction

manual, “10-9 Clock and Calendar

Function”.

Confirms and sets the time and

date. The time and date are

added to the output data.

0

No comparison

1

Comparison, excluding “near zero” when stable or overloaded

2

Comparison, including “near zero” when stable or overloaded

3

Continuous comparison, excluding “near zero”

Cp
Comparator mode

4

Continuous comparison, including “near zero”

0

Digital input, upper/lower limits

1

Weighing input, upper/lower
limits

Cp Hi, Cp lo can be selected.

2

Digital input, reference value

Cp fnc
Comparator

Cp in
Input method

3

Weighing input, reference value

Cp ref, Cp lmt can be
selected..

Factory setting

Note: “Digit” is a unit of minimum weighing value

Advertising