9-4 Description of the Class “Environment, Display”

Condition ( Cond

)

Cond 0

This parameter is for sensitive response to the fluctuation of a weight value.

Used for powder target weighing, weighing a very light sample or when
quick response weighing is required.

After setting, the balance displays FAST.

Cond 2

This parameter is for stable weighing with slow response. Used to prevent a
weight value from drifting due to vibration or drafts.

After setting, the balance displays SLOW.

Notes

In automatic response adjustment, the response rate is selected

automatically.

With “Hold function ( Hold )” set to “ON (1 )”, this item is used to set

the averaging time.

Stability band width ( 5t-b )

This item controls the width to regard a weight value as a stable value. When the fluctuation per
second is less than the parameter, the balance displays the stabilization indicator and outputs or
stores the data. The parameter influences the “Auto print mode”

5t-b 0

This parameter is for sensitive response of the stabilization indicator. Used

for exact weighing.

5t-b 2

This parameter ignores slight fluctuation of a weight value. Used to prevent

a weight value from drifting due to vibration or drafts.

Note
With “Hold function (
Hold )” set to “ON (1 )”, this item is used to set
the stabilization range.

Zero upon power-on ( p-tr )

When a hopper is attached to the weighing pan and loss-in weighing is performed, the remaining
amount of the material will become unknown if tare is performed each time a weighing starts.
When " p-tr " is set to “ 1 ”, tare is not performed at weighing start. So, the remaining amount of the
material can be monitored, when the power is turned on again after it was turned off.

Span calibration ( 5pn )

When a hopper is attached to the weighing pan and calibration is to be performed with the hopper
attached, set " 5pn " to “ 1 ”. When the tare value (hopper and other devices attached) is within the
value in the table below, calibration using the internal mass is possible.

Model Tare

value

MC-1000

500 g or less

MC-6100

5 kg or less