14. EXTENDED FUNCTION

The MC series balance has several extended functions equipped for special applications or to
troubleshoot when using the standard functions.

When the "Filter ( f1l)” is set to “ 0 “, extra items are available as shown below.

Class

Item and Parameter

Description

0
1
2
3

4

5
6

f1-b

Averaging range for

the first moving

average

7

Small


Large

When the fluctuation of a weight value is
within the averaging range, the averaging
operation starts to stabilize the value.
When the fluctuation is small, for example,
when weighing or filling a small amount of
sample, the averaging operation is always
performed and the response rate may be
slow. Under such a condition, change the
parameter.
Refer to “Averaging range ( f1-b ) and
averaging time ( f1-t )”

0 No averaging
1 0.5 second

2

1 second

3 1.5 seconds
4 2.0 seconds
5 2.5 seconds
6 3.2 seconds
7 4.8 seconds

f1-t

Averaging time for the

first moving average

8 6.6 seconds

When the fluctuation of a weight value is
within the averaging range, the averaging
operation starts. Once the duration of
averaging operations reaches the averaging
time, moving averaging starts.
This parameter sets the time to start moving
averaging.
Refer to “Averaging range ( f1-b ) and
averaging time ( f1-t )”

0

1

2
3
4
5

f2-b

Averaging range for

the second moving

average

6

Small


Large

Refer to “Filter depending on differences in

the amount to deliver powdery and liquid

material”.

0 No averaging

1 0.5 second

2 1 second

3 1.5 seconds

4 2.0 seconds

5

2.5 seconds

erfnc

Extended

function

f2-t

Averaging time for the

second moving

average

6 3.2 seconds

Advertising
This manual is related to the following products: