2 glp report – Rice Lake MC Series Mass Comparator, A&D Weighing - MC-6100 User Manual

Page 42

Advertising
background image

40

10-2 GLP Report

Set the following parameters to output the report.

z

To print the report, set the “GLP output ( info )” parameter to “1 ”, the “Data output pause

( pU5e )” parameter to “1 ”, and use MODE 3 of the AD-8121B. For details on using the printer,
refer to “12-1 Connection to the AD-8121B Printer”.

z

To output the report to a personal computer using the RS-232C interface, set the “GLP output

( info )” parameter to “ 2 ”.

z

If the time and date are not correct, set the correct time and date in “Clock ( Cl adj )” of the

function table.

Calibration report using the internal mass

When the setting is “info 1 ”:

When the setting is “info 2 ”:

~~~~~~~~~~~

A

~

&

~

D < T E R M >

MO DE L

~~~~

MC-1 00 0< T E RM >

S/ N

~~~~~

012 34 567 < TE R M >

ID

~~~~~~~

AB CDE FG < T E R M >

DA TE < T E R M >

~~~~~~

2 00 9/12 /31< T E RM >

TI ME < T E R M >

~~~~~~~~

1 2:3 4: 56< T E RM >

CA LI BRA TE D(IN T. )< T E R M>
SI GN ATU RE < T E R M >
<T E R M>
<T E R M>
-- - - - - - - - -- - - - - - < T ER M >
<T E R M>
<T E R M>

A & D
MODEL MC-1000
S/N 01234567
ID ABCDEFG
DATE 2009/12/31
TIME 12:34:56
CALIBRATED(INT.)
SIGNATURE

- - - - - - - -

Manufacturer

Model

Serial number

ID number

Date

Time

Calibration type

Signature

AD-8121 format

General data format

<TERM>

CR LF

CR

Space, ASCII 20h
Terminator, , or
Carriage return, ASCII 0Dh
Line feed, ASCII 0Ah

CR

LF