Rice Lake MC Series Mass Comparator, A&D Weighing - MC-6100 User Manual
Page 32
30
Class
Item and Parameter
Description
Cp Hi
Upper limit
Cp lo
Lower limit
Displays when Cp in 0 or
Cp in 1 is selected.
Cp ref
Reference value
Cp lmt
Tolerance
Refer to the GX series
instruction manual, “10-10
Comparator Function”.
Displays when Cp in 2 or
Cp in 3 is selected.
0
Key mode (when stable)
Accepts the PRINT key only when
the display is stable.
1
Auto print mode A
(Reference = zero)
2
Auto print mode B
(Reference = last stable value)
Outputs data when the display is
stable and conditions of ap-p,
ap-b and the reference value are
met.
prt
Data output mode
3
Stream mode /
Interval memory mode
With data 0, outputs data
continuously; with data 2, uses
interval memory.
0
Plus only
Displayed value>Reference
1
Minus only
Displayed value<Reference
ap-p
Auto print polarity
2
Both
Regardless of displayed value
0
10 digits
1
100 digits
ap-b
Auto print difference
2
1000 digits
Difference between reference
value and displayed value
0
Not used
1
Stores unit mass in counting mode
2
Stores weighing data
data
Data memory
3
Stores calibration data
Related items:
prt, int, d-no, 5-td, info
0
Every measurement
1
2 seconds
2
5 seconds
3
10 seconds
4
30 seconds
5
1 minute
6
2 minutes
7
5 minutes
int
Interval time
8
10 minutes
Interval time in the interval
memory mode when using prt 3,
data 2
0
No output
d-no
Data number output
1
Output
Refer to the GX series
instruction manual, “12. DATA
MEMORY”.
0
No output
1
Time only
2
Date only
5-td
Time/Date output
3
Time and date
Selects whether or not the time
or date is added to the weighting
data. For details, Refer to the
GX series instruction manual,
"10-9 Clock and Calendar
Function".
0
No output
5-id
ID number output
1
Output
Selects whether or not the ID
number is output.
0
No pause
dout
Data output
pU5e
Data output pause
1
Pause (1.6 seconds)
Selects the data output interval.
Factory setting
Note: “Digit” is a unit of minimum weighing value