7-5 Calibration Test Using an External Weight

This function tests the balance weighing accuracy using an external mass and outputs the result.
This is available only when the “GLP output (info )” parameter is set to “ 1 ” or “ 2 ”. (Calibration test
does not perform calibration.)

When the MC-1000 is used

1 Connect the AC adapter and warm up the balance

for at least 30 minutes with nothing on the pan.

2 Press and hold the CAL key until CCout is

displayed, then release the key.


3 The balance displays CC 0 .

z

If you want to change the target weight (a list of

usable weights is shown on page 18), press the

SAMPLE key and proceed to step 4.

z

If you use the target weight value stored in the

balance, proceed to step 5.

4 Specify the target weight value as follows:

SAMPLE key

To switch between the target

weight selection mode (All of
the digits blinking) or the
value adjustment mode (The
selected digits blinking).
With the MC-1000, the digits in
the 2

nd

, 3

rd

, and 4

th

decimal

places can be adjusted.
With the MC-6100, the digits in
the 2

nd

and 3

rd

decimal places

can be adjusted.

RE-ZERO key

To select the target weight

(All of the digits blinking) or
adjust the value (The
selected digits blinking).
Refer to page 18

PRINT

key

To store the new weight

value. Even if the AC
adapter is removed, the data
is maintained in non-volatile
memory.

CAL key

To cancel the operation and

return to

1

CC 0 .



Press and

hold the key

Release

the key

e.g. Target

weight value
200.0123 g

Select a

weight

To next page