4 calibration using an external weight – Rice Lake MC Series Mass Comparator, A&D Weighing - MC-6100 User Manual

Page 22

Advertising
background image

20

7-4 Calibration Using an External Weight

This function calibrates the balance using an external weight.


When the MC-1000 is used

1 Connect the AC adapter and warm up the balance

for at least 30 minutes with nothing on the pan.

2 Press and hold the CAL key until Calout is

displayed, then release the key.

3 The balance displays Cal 0 .

z

If you want to change the calibration weight (a

list of usable weights is shown on page 18),
press the SAMPLE key and proceed to step 4.

z

If you use the calibration weight value stored in

the balance, proceed to step 5.

4 Specify the calibration weight value as follows:

SAMPLE

key

To switch between the

calibration weight selection
mode (All of the digits blinking)
or the value adjustment mode
(The selected digits blinking).
With the MC-1000, the digits in
the 2

nd

, 3

rd

, and 4

th

decimal

places can be adjusted.
With the MC-6100, the digits in
the 2

nd

and 3

rd

decimal places

can be adjusted.

RE-ZERO

key To select the calibration

weight (All of the digits
blinking) or adjust the value
(The selected digits blinking).
Refer to page 18

PRINT

key

To store the new weight

value. Even if the AC
adapter is removed, the data
is maintained in non-volatile
memory.

CAL key

To cancel the operation and

return to

1

Cal 0 .

To next page

Select a

weight

Press and

hold the key

Release

the key

e.g. Calibration

weight value
200.0123 g

Press