36

KF format

5if type 2

This is the Karl-Fischer moisture meter format and is used when the peripheral equipment can only
communicate using this format.

z

This format consists of fourteen characters excluding the terminator.

z

This format has no header characters.

z

The polarity sign is placed before the data, with spaces in place of leading zeros, if the data is

not zero or overloaded.

z

This format outputs the unit only for a stable value.

MT format

5if type 3

z

A header of two characters indicates the balance condition.

z

The polarity sign is used only for negative data.

z

The weighing data uses spaces in place of the leading zeros.

z

The character length of this format changes dependent upon the unit

NU (Numerical) format

5if type 4

This format outputs only numerical data.

z

This format consists of nine or ten characters excluding the terminator.

With the MC-1000, when the numerical data excluding the decimal point exceeds eight
characters, the data format will be ten characters long.

z

The polarity sign is placed before the data with the leading zeros. If the data is zero, the plus

sign is used.

(Data example exceeding 8 characters)







+ 1 0 0 0 . 0 0 0 0 g

CR LF

Data

Unit

Terminator

S

1 0 0 0 . 0 0 0 0 g

CR LF

Header

Unit Terminator

Data

+ 0 0 0 . 0 1 2 7

CR LF

Data

Terminator

+ 1 0 0 0 . 0 0 0 0

CR LF

Data

Terminator

g

Stable value

Unstable value