2 manual response adjustment – Rice Lake MC Series Mass Comparator, A&D Weighing - MC-6100 User Manual

Page 19

Advertising
background image

17

Notes

If improper performance is found in the self check, the

balance displays CH no . Contact the local A&D
dealer for repair.

If the automatic response adjustment fails, the

balance displays CH ng . Check the ambient
conditions such as breeze and vibration, also check
the weighing pan. Then, perform the adjustment again.
To return to the weighing mode, press the CAL key.

If the automatic response adjustment is awkward, try

to refine it using the manual response adjustment.

6-2 Manual Response Adjustment

This function manually updates the response adjustment.

Operation

1 Press and hold the MODE key until RESPONSE is

displayed, then release the key.

Press the MODE key again quickly.

2 Press the MODE key to select a rate of the response

adjustment. Either FAST, MID. or SLOW can be
selected.

3 After a few seconds of inactivity the balance displays

e

end . Then, it returns to the weighing mode and

displays the updated response indicator. The

response indicator remains displayed for about 30
seconds.

Note
The response adjustment can be changed at “Condition
(
Cond)” of “Environment, Display ( ba5fnc )” in the
function table. For details, refer to “9. FUNCTION
TABLE”.





Press and hold

the key

Each pressing

switches

the indicators

Release the key

and press again

After a while