Rice Lake MC Series Mass Comparator, A&D Weighing - MC-6100 User Manual

Page 55

Advertising
background image

53

Filter depending on differences in the amount to deliver powdery and liquid
material


z

When the weight value is not stable even without load fluctuations

Increase the averaging range for the first moving average parameter ( f1-b ).
Increase the averaging time for the first moving average parameter ( f1-t ).
Strengthen the digital filter. (Increase the function setting “Condition ( Cond )” parameter.)

z

When response is slow during full and medium flow

Decrease the averaging range for the first moving average parameter ( f1-b ).
Weaken the digital filter. (Decrease the function setting “Condition ( Cond )” parameter.)

z

When response is slow during dribble flow

Decrease the averaging time for the first moving average parameter (f1-t).
Weaken the digital filter. (Decrease the function setting “Condition ( Cond )” parameter.

z

To increase stability without load fluctuations

Increase the averaging range for the second moving average parameter ( f2-b ).
Increase the averaging time for the second moving average parameter ( f2-t ).





















Advertising