2 acknowledge code and error codes – Rice Lake MC Series Mass Comparator, A&D Weighing - MC-6100 User Manual

Page 51

Advertising
background image

49

13-2 Acknowledge Code and Error Codes

When the “Serial interface function ( 5if )” parameter is set to “ erCd 1 ”, the balance outputs <AK>

code or error code to each command as follows:

<AK> (06h)

Acknowledge in ASCII code.

z

When the balance receives a command to request data and can not process it, the balance

transmits an error code (EC, Exx).

When the balance receives a command to request data and can process it, the balance

outputs the data.

z

When the balance receives a command to control the balance and can not process it, the

balance transmits an error code (EC, Exx).

When the balance receives a command to control the balance and can process it, the balance

transmits the acknowledge code.

Among commands to control the balance, the following transmit the acknowledge code both

when the balance receives the command and when the balance has accomplished the
command. If the command can not be processed properly, the balance transmits an error code
(EC, Exx). This error can be released using the CAL command.

CAL command (Calibration command)

ON command (Display ON command)

P command (Display ON/OFF command)

R command (RE-ZERO command)

R command


z

When a communication error has occurred due to external noise, or a parity error has

occurred due to transmission error, the balance transmits an error code. In this case, send the
command again.


Command received

Command completed

Weighing mode

Zero display

In process

Personal computer Time Balance

R command

R

CR LF

AKCR LF

AKCR LF

Advertising
This manual is related to the following products: