Calibration, 1 calibration group, Caution – Rice Lake MC Series Mass Comparator, A&D Weighing - MC-6100 User Manual

Page 20

Advertising
background image

18

7. CALIBRATION

7-1 Calibration Group

Calibration

z

Automatic self calibration (calibration due to changes in temperature)

z

Calibration using the internal mass (one-touch calibration)

z

Calibration using an external weight

Calibration test

z

Calibration test using an external weight (Calibration test does not perform

calibration.)

Correction of the internal mass value

Terms

The following terms are defined as follows:
Internal mass

= Built-in calibration weight

External weight = A weight that you have. Referred to as a calibration weight when used for calibration.
Calibration weight = A weight used for calibration
Target weight

= An external weight used for calibration test

Caution

z

Calibration adjusts the balance for accurate weighing.

Besides periodic calibration and before each use, perform calibration when:
• the balance is installed for the first time.
• the balance has been moved.
• the ambient environment has changed.
z

Do not allow vibration or drafts to affect the balance during calibration.

z

To output the data for GLP using the RS-232C interface, set “GLP output ( info )” of “Data

output ( dout )”. For details, refer to “9. FUNCTION TABLE”. Time and date are added to the
GLP report. If the time or date is not correct, adjust them. For details, refer to the GX series
instruction manual, “10-9 Clock and Calendar Function”.

z

Calibration test is available only when “GLP output ( info )” of “Data output ( dout )” is set to

“1 ” or “ 2 ”.

z

The calibration and calibration test data can be stored in memory. To store them, set “Data

memory ( data

)” to “ 3 ”. For details, refer to the GX series instruction manual, “12. DATA

MEMORY”.

Caution on using an external weight

z

The accuracy of an external weight can influence the accuracy of weighing. Select an

appropriate weight as listed below:

Model

Usable calibration weight

Adjustable range

MC-1000

1000g, 900g, 800 g, 700 g, 600 g, 500 g, 400 g,
300 g, 200 g

-0.0150 g to +0.0159 g

MC-6100 6000

g,

5000 g, 4000 g, 3000 g, 2000 g

-0.150 g to +0.159 g

The calibration weight in bold type: factory setting
The calibration weight value can be adjusted within the range above.

Display

z

This indicator means “the balance is measuring calibration data”. Do
not allow vibration or drafts to affect the balance while this indicator is
displayed.