Rice Lake MC Series Mass Comparator, A&D Weighing - MC-6100 User Manual

Page 39

Advertising
background image

37

CSV format

5if type 5

z

Separates the data of A&D standard format and the unit by a comma (,).

z

Outputs the unit even when the data is overloaded.

z

When the ID number, data number, time and date are added, outputs the ID number, data

number, date, time and weighing data in this order and separates each item by a comma and
treats all the items as one group of data.


Note
To add the ID number, data number, time and date, the function settings must be changed.

9-7 Description of the Data Format Added to the Weighing Data

ID number

dout 5-id 1

The number to identify a specific balance.

z

This format consists of seven characters excluding the terminator.

Data number dout d-no 1

This format outputs the data number just before the data is transmitted using the RS-232C interface.

z

This format consists of six characters excluding the terminator.

z

When CSV format ( 5if type 5 ) is selected, the period (

. ) is replaced with a comma (

,

).

Note
The data number is added only to the weighing data that is stored in memory.

Date dout 5-td 2

or

3

z

The date output order can be changed in “Clock ( Cl adj )”. Outputs the year in four-digit

format.


Time dout 5-td 1

or

3

z

Outputs time in 24-hour format.

Note

When the data described above is added to the weighing data, the output is in the following

order: ID number, Data number, Date, Time and Weighing data.

2 0 0 9 / 1 2 / 3 1

CR LF

CR LF

O L , +

S T , + 1 0 0 0 . 0 0 0 0 ,

g

CR LF

LAB-123, No,012, 2009/12/31, 12:34:56, ST,+1000.0000, g<CR><LF>

ID number Data number Date

Time

Weighing data

9 9 9 9 9 9 9 E + 1 9 ,

g

Advertising