Symptom

Probable Cause

Action

The weight reading
on the indicator is

unstable.

a) The circuit board in

the control panel

may be wet. If there
is a junction box for
the load cells on

your system, check
there for moisture
as well.

b) A load cell cable

may be pinched or

damaged.

a) Dry any areas that are

contaminated with moisture.

Check for leaks and reseal.

b) Contact Rice Lake Weighing

Systems or a qualified dealer

for support. Cutting the load
cell cable will void the
warranty. Special repair
techniques are required.



on display

Will occur when start is
pressed

a) On some systems

prior to 1999 when
the scales gross

weight is zero or
negative.

b) On any system if

start is pressed
when the motion
light is on.

a) Make sure the scale will have

a positive weight.

b) Wait for scale to settle.

The scale has a
positive error when
loading or a negative

error when
unloading

a) Mechanical binding

problem on scale

a) Check all hydraulic hoses and

conduit, they must be long
and loose enough not to

exert a force on the scale.
Check all load cell covers to
see that they are not bent or

contacting the lift arm or
cylinder.