Setup procedures, Display error codes, Err 33 – Hardy HI 2151/30WC Single-Scale Controller User Manual

Page 14: Err 34, Err 52, Blind unit operation setup, Setup procedures -2 display error codes -2, Err 33 -2 err 34 -2 err 52 -2, Blind unit operation setup -2

Advertising
background image

HI 2151 SERIES WEIGHT CONTROLLERS REMOTE I/O OPTION

3-2

Setup Procedures

Step 1.

Enter the Option Menu by pressing the 7/Option button. (Dis-
play shows the first option available)

Step 2.

Press the up arrow until RIO is displayed on the screen.

Step 3.

Press the Enter button two times. (Display shows the cur-
rently selected Baud Rate value)

Step 4.

Use the up or down arrow to select a baud rate 57600, 115200
or 230400. (Display shows the currently selected value)

Step 5.

Press the Enter button two times.

Step 6.

If a change is necessary, press the Test/Clr button.

Step 7.

Use the numeric buttons and enter the PLC rack number.
(Maximum 63)

NOTE:

The rack number is displayed in decimal on the weight controller, and octal in the
PLC. You cannot use 0 for the PLC rack number.

Step 8.

Press the Enter button.

Step 9.

Press the Enter button to see the quarter number. (Display
shows the currently selected value)

Step 10. If a change is necessary, press the Test/Clr button.
Step 11. Use the numeric buttons and enter the PLC quarter number

(maximum 3).

NOTE:

The quarter rack number in the PLC is displayed in decimal. Qtr 0 = PLC Group 0,
Qtr 1 = PLC Group 2, Qtr 2 = PLC Group 4, Qtr 3 = PLC Group 6.

Step 12. Press the Enter button.
Step 13. Press the Enter button to view last quarter status.
Step 14. Use the up or down arrow buttons to select Yes or No to indi-

cate whether or not this is the last quarter of this rack cur-
rently in use.

Step 15. Press the Enter button.
Step 16. Press the Exit button.
Step 17. Press the Exit button.

NOTE:

If any data was changed a Reboot is required.

Step 18. Now power-down the instrument and re-apply power to have

new menu selections activated. In addition, you must perform
a manual or auto configuration of the PLC.

Display Error Codes

These display error codes are in addition to those listed in the HI
2151WC manual.

ERR 33

Invalid quarter number entered. Select a value from 0 - 3.

ERR 34

Invalid rack number entered. Select a value from 1 - 63.

ERR 52

Too many serial ports are installed.

Blind Unit Operation