Chapter 5 - Block Transfers

5-9

Block Read
Command Number
7: BCD Output
Configuration

TABLE 5-10: BLOCK READ COMMAND NUMBER 7: BCD OUTPUT CONFIGURATION

Block Read
Command Number
8: Configuration of
Analog Output

BLOCK READ COMMAND NUMBER 7: BCD Output Configuration

START

WORD DEFINITIONS:

#WORDS

WORD

Command number: A value of 7 (decimal)

bit 0

1

bit 1

1

bit 2

1

bit 3

0

bit 4

0

bit 5

0

bit 6

0

bit 7

0

Format of output

bit 8

If set, will update BCD output when “print” button or remote
function is activated

bit 9

Reserved for future use

bit 10

If set, will output weight data currently displayed

bit 11

If set, will output tare value

bit 12

If set, will output net weight

bit 13

If set, will output gross weight

bit 14-15

Reserved for future use

1

0

Reserved for future use

1

1

TOTAL NUMBER OF WORDS

2

BLOCK READ COMMAND NUMBER 8: Configuration of Analog Output

START

WORD DEFINITIONS:

#WORDS

WORD

Command number: A value of 8 (decimal)

bit 0

0

bit 1

0

bit 2

0

bit 3

1

bit 4

0

bit 5

0

bit 6

0

bit 7

0

Weight parameter to be transmitted

bits 8 - 15

A value from 0 to 4 (0 = Gross, 1 = Net, 2 = Rate-of-Change,
3 = Peak Force, 4 = Totalize amount)

1

0