Chapter 3 - Setup

3-3

About Blind Units

An HI 2151WC Weight Controller that cannot be programmed or con-
figured from the front panel is a blind unit. In a blind unit, the Remote
I/O parameters are configured using both the interior and exterior dip
switches. (See Tables below)

NOTE:

You must power-down and power up the instrument to have new switch positions acti-
vated. You must also perform a manual or auto configuration of the PLC.

Blind Unit
Configuration

Step 1.

Disconnect the power cord from the instrument.

Step 2.

Set the Interior Dip Switches. (See Table 3-1)

ON = 1
OFF = 0

NOTE:

Remember to select the appropriate jumper positions on the Remote I/O Option
Board.

INTERIOR DIP SWITCHES

Switch Position - S2 which is located
on the Power/Relay board

1

n/u

2

n/u

3

n/u

4

n/u

5

last quarter in rack
ON = Yes
OFF = No

6

Blind Unit
ON = Yes
OFF = No

7

A1 (See Binary
Baud Rate Table
3-1)

8

A0 (See Binary
Baud Rate Table
3-2)

TABLE 3-1: INTERIOR DIP SWITCHES

BINARY BAUD

RATE

A1

A0

BAUD

0

0

57.6K

0

1

115.2K

1

0

230.4K

TABLE 3-2: