Block Write
Command Number
51: Activate Scale
Functions

TABLE 5-16: BLOCK WRITE COMMAND NUMBER 51: ACTIVATE SCALE FUNCTIONS

BLOCK WRITE COMMAND NUMBER 51: Activate Scale Functions

START

WORD DEFINITIONS:

#WORDS

WORD

Command number: A value of 51 (decimal)

bit 0

1

bit 1

1

bit 2

0

bit 3

0

bit 4

1

bit 5

1

bit 6

0

bit 7

0

Remote Functions Byte

bit 8

Acquire TARE (Set tare value = current gross weight)*

bit 9

Initiates print on standard RS232 or optional BCD port*

bit 10

Add current Net weight to Total*

!

bit 11

Clear Peak Hold register*

!

bit 12

Clear Totalizer Accumulation*

!

bit 13

Zero the instrument*

bit 14

Enable Zero Tracking (Blind Unit Only)

bit 15

Reserved for future use

*Note: The bit must be toggled to activate this function
!Note: Only active if the instrument is ordered with this option

1

0

TOTAL NUMBER OF WORDS

1