Industrial Data Systems IDS 410 User Manual

Page 10

Advertising
background image

IDS 410/422 User's Manual

9

Connect IDS 410/422 to host device (optional)

The Serial Communications Port 2 (TB3) provides a continuous output of scale data. This is

used for connection to a host device such as a computer or to a remote display (scoreboard).

Continuous output transmission can be enabled or disabled and the

output format can be customized to interface with the host device (see OPERATIONS

PARAMETERS - Continuous TX2 Format, page 5-10). See the HARDWARE

INSTALLATION AND WIRING section (page 8-2) for pin assignments and sample cable

drawings.

Apply AC/DC Power to the IDS 410/422

Once you have connected the IDS 410/422 to the load cell and optional equipment, plug the

power cord into AC or DC power.

The IDS 410/422 has an onboard transformer for AC applications. Insert the power cord into the

AC socket-outlet and the unit will power up, the plug on the power supply cord is considered the

units disconect device. The socket-outlet shall be installed near the equipment and shall be easily

accessible.

For DC applications the motherboard is equipted with a 2 position terminal block marked “PWR-

DC” connect the positive source side to the terminal marked + and negative to the terminal
marked

-.

See the HARDWARE INSTALLATION AND WIRING section (page 8-1) for pin

assignments and a sample cable drawing.

On power-up, the IDS 410/422 performs the following tests:

1 A/D Conversion Test - the IDS 410/422 checks the A/D converter to verify that data

conversions are taking place.

2 Keyboard Scan - the keyboard is scanned to verify that no keys are engaged..

3 Memory Test - the IDS 410/422 displays "rtESt" while it performs a memory test on internal

memory. If a problem is encountered during the test, an error message is displayed.

4 Lamp Test - all segments on the numeric display and all LED's are turned on for 5 seconds.

5 Conversion Display - the IDS 410/422 displays "Ad xxx", where "xxx" is the conversion

count. This test takes approximately 10 seconds.

6 Stability Check - the IDS 410/422 displays "Cxxxx" where "xxxx" is the raw data difference

in load cell readings. The IDS 410/422 waits until the data difference is within a stable

range.To exit the above power on test functions, press the

clear [ZERO] key. This is possible

only if the IDS 410/422 has not been locked. The IDS 410/422 is now ready for initialization,

calibration, and configuration.