IDS 410/422 User's Manual

52

Function 74: Configure Timers

The display prompts “tr” when function 74 is activated. Enter the timer number to be modified

(1-5) then press the ENTER key. The first parameter (L) is the time length in tenths of a second.

The second parameter (Fn) is the function number to be executed when the timer times out.

Timer #5 reloads after it counts down to zero. This provides more accurate timing of repetitive

events ( such as pulse output, speed calculations and interval timing, etc.) See timer events in the

Scale Basic programming guide.

Parameter 76: Debug Mode for Event Monitors and Scale Basic

The debug mode is used to debug event monitors and scale basic functions. Enter a 1 to activate

debug mode, enter a 0 to diable debug mode. When debug mode is enabled Scale Basic functions

single step. The subroutine stack level and the function number are displayed when entering a

function. EXAMPLE: “0 F12”. Press the ENTER key to begin execution of the function. The

instruction address and instruction code is displayed. EXAMPLE: “21 255”. Continue to press

the ENTER key to step through the program. NOTE: the first step displayed for all Scale Basic

funtions is a GOTO (244).

Function 78: Initialize ID Memory

Use function 78 to initialize ID memory after it has been corrupted. The ID memory can become

corrupted when power is lost (when no battery is installed) or when the battery is allowed to

discharge below 2 volts.

Parameter 79: Enable Multi-Point Linearization

Enter a 1 to enable multi-point linearization, enter a 0 to disable. The IDS 410/422 uses the

calibration factors set in parameter 80 below when multi-point linearization is enabled.