IDS 410/422 User's Manual

29

Parameter 17: Full Scale Graduations (10,000)

Enter the number of graduations at full scale. The full-scale graduations are calculated

using the count-by, decimal position, and full-scale weight.

Full Scale Graduations = Full Scale Weight / count-by .

For example, if Full Scale Weight = 600.0 and count-by = 0.2 then Full Scale Graduations

= 600.0 / 0.2 = 3000.

Parameter 18: Overload Graduations (10,200)

Enter the number of graduations at which the scale is in overload status. The IDS 410/422

displays Err oL on the numeric display when the scale weight exceeds overload. For

example, if the Full Scale Graduations is 10,000 grads and the scale is rated for 2%

overload, then Overload Graduations = 10,000 + (10,000 x 0.02) = 10,200.

Parameter 19: Deadload Offset (1)

This parameter is automatically calculated by the calibration function

.

Parameter 20: Disable Power-On Zero (1 [disabled])

Enter a 1 to disable the power-on zero function. Enter a 0 to enable power-on zero. The

power-on zero function attempts to acquire zero when the power on tests have completed.

If the weight on the scale is above the zero range the display prompts "Err 0". Press the

ENTER key to view the scale weight. Remove the weight from the scale to enter weighing

mode.

Parameter 21: Deadload Factor

The deadload factor is automatically calculated by the calibration function. This parameter

is not accessible from the keyboard.

Parameter 22: Weight Conversion Factor

The weight conversion factor is automatically calculated by the calibration function. This

parameter is not accessible from the keyboard.