Error Messages

Error No. Description

Err 1

Keyboard error. Occurs during power up while a key on the keyboard is being

pressed or one of the remote switch inputs is activated.

Err 2

Restart Trap. The microprocessor accessed nonexistent memory location.

Usual cause is A/C power "glitch" or static electric discharge.

Err 3

Watchdog Timeout. The weight display has not been updated within the

watchdog time-out period. Usual cause is A/C power glitch or static electric

discharge.

Err 4

Battery Error. (Optional) battery voltage level was below 2 V when the

410/422 was powered off.

Err 5

Earom memory error. A checksum error has occurred when reading the

configuration earom memory. Check calibration parameters and rewrite the

earom memory. If error continues, replace earom (U12).

Err 5

Earom Timeout, unable to write to the earom memory.

Err 6

Ram memory error. A ram memory error has occurred. Replace U5 if error

continues.

Err 7

The A/D converter is "locked up". Check load cell wiring.

Err 8

Negative deadload error.

Err 9

Count-by error. The entered value is outside the scale count-by set during

calibration.

Err 10

Push button zero error.

Err 11

A to D convertor not working. Check load cell wiring.

Err 12

Negative deadload. Check loadcell wiring. Enter a deadload offset in

configuration parameter 19.

Err 14

Page Format line length error. The page format output exceeded the maximum

line length.

Err 15

The function selected is locked. Switch S1 on the IDS 410/422's circuit board

locks and unlocks configuration parameters.

Err 16

Scale Basic Stack Overflow. Too many setpoints in user function running

simutainuously.

Err 18

Referenced a Function Instruction that does not exits.

Err 21

ID not found.

Err 22

Scale Basic Prompt error.

Err 23

ID memory full.

Err 24

Link list error.

Error OL Calibration error or scale is over-loaded (above scale capacity).

undEr

Calibration error or scale is under-loaded (below readable weight).

SuSPEn Scale Basic execuded a RESUME instruction without a corisponding

SUSPEND instruction.