IDS 410/422 User's Manual

37

Function 59: Display Calibration Audit Number

Use function 59 to view the calibration audit number , the configuration number, the software ID

number, and the status of the internal function lock.

The IDS 410/422 provides a calibration audit number and a configuration number. They are

used to indicate changes to calibration data and configuration data and can be used in place of

mechanical seals by some regulatory agencies. The IDS 410/422 audit number increments each

time a calibration parameter is changed (deadload or weight conversion factor). The

configuration number increments each time a password level 3 parameter is changed. If physical

seals are used, then place the internal function lock in the locked position before sealing.

Calibration and password level 3 parameters can not be accessed when the function lock is in the

locked position.

1 Hold the GROSS/NET key down and press the UNITS key to enter configure mode.

2 Enter CFG 59. The IDS 410/422 prompts "A xxx" where xxx is the current audit number.

3 Press the ENTER key. The IDS 410/422 prompts "C xxxxx" where xxxxx is the current

calibration number.

4 Press the ENTER key. The IDS 410/422 prompts "xxxxx" where xxxxx is the ID number of

the software installed in the IDS 410/422.

5 Press the ENTER key. The IDS 410/422 prompts "un-LoC" if the function lock is in the

unlocked position, or "Loc Ed" if the function is in the locked position.

Function 60: Calibrate Scale - Deadload First

Use function 60 to calibrate the scale by reading the deadload first and then reading the span. Set

the Scale Operation Parameters before using function 60. Best results are obtained when the

IDS 410/422 and load cells have reached thermal equilibrium (power applied for approximately

30 to 120 minutes depending on load cell size). NOTE: function 60 automatically adjusts

parameter 19 (internal A/D deadload factor).

1 Hold the GROSS/NET key down and press the UNITS key to enter configure mode.

2 Enter CFG 60. The IDS 410/422 prompts "dEAdld". Remove all weight from the scale.

Press the ENTER key. The IDS 410/422 reads the scale's deadload weight.

3 The IDS 410/422 prompts "CALib". Place the calibration weight on the scale. Enter the

calibration weight using the

item, +, - keys. Press the ENTER key. The IDS 410/422 reads the

calibration weight.

4 The IDS 410/422 prompts "CFG 60". Press the CLEAR key to exit the configure mode.