IDS 410/422 User's Manual

34

Parameters 35, 36, 37, and 38: Print Codes

The IDS 410/422 provides 4 programmable print codes. The print codes are used to send special

setup parameters to the printer, such as select print size or select print font. Each print code is up

to 4 characters long. Consult your printer manual for the print codes that you may want to use.

Enter the codes in decimal format (an ASCII carriage return = 13). Select parameter 35 to begin

print code 1 entry. The IDS 410/422 prompts "C0 XXX" where xxx is the current character 0

print code value. Enter data for character 0. The IDS 410/422 prompts "C1 XXX". Continue

the code entry. Enter a 0 to end print code entry.

Parameter 39: Enable TTL Remote Control (1 [enabled])

Use parameter 39 to enable remote control using TTL inputs 1 through 3 (TB3 terminals 8, 9

and 10).

Enter a 1 to activate scale basic functions 1 - 3. For example, grounding TB3 pin 8 (TTL input 1)

activates scale basic function 1.Enter a 0 to disable TTL remote control.

Display Intensity, Watch Dog Timer

Parameter 40: Display Intensity (10)

Use parameter 40 to set the brightness of the IDS 410/422's numeric display and LED indicators.

The brightness can be set from 1 (dimmest) to 15 (brightest).

Parameter 41: Battery backup (0[off])

Parameter 41 is used to preserve the time and date, memory registers, and ID memory

when power is lost. Set parameter 41 to ON (1) this will prevent the IDS 410/422 from

clearing the memory when power is restored. Set parameter 41 to OFF (0) to erase

memory when power is restored.

Parameter 42: Enable Watch Dog Timer (0 [no])

The watch dog timer resets the IDS 410/422 when the weight display is not updated within the

time-out time (0.2 sec). Set the Enable ON (1) to reset after time-out. Set the Enable OFF (0) to

turn off the watch dog timer.

Parameters 43-50: Fixed Registers

The Fixed Registers are used by the Trip-Level Event Monitor to determine when a trip-level is

reached. These parameters are set by the Mode of Operation selection (parameter 73) as needed.