6.6 Mains connection

Power is supplied via the external mains adapter. The stated voltage value must be
the same as the local voltage.
Only use original KERN mains adapters. Using other makes requires consent by
KERN.

6.7 Rechargeable battery operation (optional)

The rechargeable battery is charged via the delivered power cable.
Before the first use, the rechargeable battery should be charged by connecting it to
the mains power cable for at least 15 hours. The operation time of the rechargeable
battery is approx. 70 hours. When connecting a second balance, the operation time
will be reduced. The charging period for total charge is c. 12 hours
To save the rechargeable battery, in the menu (see chap.12.2) the automatic
switch-

off function [„

“  „

„] can be activated, switch-off time selectable

after 0, 3, 5, 15, 30 minutes.

If an arrow appears on the weight display  above the battery symbol

or

"bat lo" when turning on the balance, this is an indication that the capacity of the
rechargeable battery will soon be exhausted. The balance will be ready to operate
for about another 10 hrs., then it will switch off automatically. Connect the power
cable as soon as possible to load the rechargeable battery.

The LED display informs you during loading about the loading status of the
rechargeable battery.

red:

Voltage has dropped below prescribed minimum. Connect mains
adapter to load the rechargeable battery.

green:

Battery is completely discharged

yellow: Battery very low. Connect mains adapter as soon as possible to load the

rechargeable battery.

6.8 Connection of peripheral devices

Before connecting or disconnecting of additional devices (printer, PC) to the data
interface, always disconnect the balance from the power supply.
With your balance, only use accessories and peripheral devices by KERN, as they
are ideally tuned to your balance.

6.9 Initial Commissioning

In order to obtain exact results with the electronic balances, your balance must
have reached the operating temperature (see warming up time chap. 1).
During this warming up time the balance must be connected to the power supply
(mains, accumulator or battery).
The accuracy of the balance depends on the local acceleration of gravity.
Strictly observe hints in chapter Adjustment.