7.4 Adjustment bulk material scales CFS 50K-3 models

Not pre-configured by KERN

Operation

Display

Switch-on balance and during the selftest press

.

Use the numeric keys to enter password:

Default password "0000"

Confirm input by

.

Use

to select bulk material scales („ Remote“).

The appeared [] displays the current balance.

Acknowledge with

.

If necessary, at balance zero display using

select

the weighing unit [kg or lb], which shall be used for
adjustment. The appeared [] displays the current
weighing unit.

Acknowledge with

.

Ensure that there are no objects on the weighing pan.

After successful standstill control press the

button.

Enter value of the necessary adjustment weight using

numeric keyboard and acknowledge with

.

In order to achieve high-quality weighing results in the
sense of the measuring technology, it is recommended
to select the nominal value as high as possible.
We recommend 80 % max.

Place the required adjusting weight carefully in the

centre of the weighing plate and press

to confirm.

After the adjustment the balance will carry out a self-test.
Remove adjusting weight during selftest, balance will
return into weighing mode automatically.
In case of an adjustment error or incorrect adjusting
weight the display will show an error message
(

), repeat adjustment process.