Instant help – KERN & SOHN CCS Version 2.0 User Manual

Page 80

Advertising
background image

80

CFS/CCS-BA-e-1220

18 Instant help

In case of an error in the program process, briefly turn off the balance and
disconnect from power supply. The weighing process must then be restarted from
the beginning.

Fault

Possible cause

The displayed weight does
not glow.

The balance is not switched on.

The mains supply connection has been interrupted

(mains cable not plugged in/faulty).

Power supply interrupted.

(Rechargeable) batteries are inserted incorrectly or

empty

No (rechargeable) batteries inserted.

The displayed weight is
permanently changing

Draught/air movement

Table/floor vibrations

Weighing pan has contact with other objects.

Electromagnetic fields / static charging (choose

different location/switch off interfering device if
possible)

The weighing result is
obviously incorrect

The display of the balance is not at zero

Adjustment is no longer correct.

Great fluctuations in temperature.

Warm-up time was ignored.

Electromagnetic fields / static charging (choose

different location/switch off interfering device if
possible)