8 Linearisation


Linearity shows the greatest deviation of a weight display on the scale to the value
of the respective test weight according to plus and minus over the entire weighing
range.
If linearity deviation is discovered during a testing instrument control, you can
improve this by means of linearization.

Carrying out linearization is restricted to specialist staff possessing well

acquainted with the workings of weighing scales.

The test weights to be use

d must be adapted to the weighing scale’s

specifications; see chapter 3.4 “testing instruments control”.

Observe stable environmental conditions. Stabilisation requires a certain

warm-up time.

After successful linearisation you will have to carry out calibration;

see chapter 3.4 “testing instruments control”.

8.1 Linearisation CFS 3K-5, CFS 6K0.1, CFS 15K0.2, CFS 30K0.5 models:

Operation

Display

 Switch-on balance and during the selftest press

.

With the numerical keys input the password „ 9999“:

Confirm with

.

„tECH LoCAL“ is displayed, should it not be

displayed, with

select „tECH LoCAL“.

Acknowledge with

.

The appeared [] displays the current balance.