9.4 Weighing with tare

A tare value can be entered for the reference as well as for the bulk material
balance. Before setting a tare values select active balance, s. chap. 9.3.

9.4.1 Taring

 Deposit weighing vessel. After successful stop check press the TARE button.

The zero display and the symbol [] above NET will appear.
The weight of the container is now internally saved.

 Weigh the material, the net weight will be indicated.
 After removing the weighing container, the weight of the weighing container

appears as negative display.

 To delete the tare value, remove load from weighing pan and press the TARE-

key.

 The tare procedure can be repeated as many times as necessary, for example

with initial weighing of several components for a mix (add-on weighing). The
limit is reached when the total weighing range capacity is full.

9.4.2 Numeric entering of tare weight

 Unload and reset to zero the balance

 Input known tare weight numerically with decimal point and press key TARE.

The entered weight will be stored as tare weight and displayed with negative
sign. The [] symbol over NET appears.

 Put the filled weighing container on the balance, the net weight will be

displayed.

 The tare value remains stored until it will be deleted by TARE.

The tare value will be rounded up according to the readability of the balance,
e.g. at a balance 60 kg max/5 g readability the input value of 103 g will be
displayed as -105 g.

Advertising